Calibration Scope Search
Scope of Calibration Capabilities
Access SIMCO’s calibration capabilities by parameter, range, instrument type, or measurement discipline. Quickly identify whether our accredited capabilities align with your application requirements.
Parameter
Range
CMC (±)
Comments
Contact
Absolute RF Power – Measure
0 dBm at 50 MHz
0.6 %
Power meter & thermistor with DC substitution
Absolute RF Power – Measure – (-20 to +20) dBm
50 MHz to 3 GHz
(3 to 26.5) GHz
100 kHz to 50 MHz
> 50 MHz to 2 GHz
(> 2 to 12.4) GHz
(> 12.4 to 18) GHz
0.088 dB
0.13 dB
1.8 %
1.6 %
1.8 %
1.9 %
Agilent/HP 8902A w/8485A
Agilent/HP 8902A w/8485A
E4418B power meter w/8482A sensor
E4418B power meter w/8481A sensor
E4418B power meter w/8481A sensor
E4418B power meter w/8481A sensor
Absolute RF Power – Measure – 0 dBm
50 MHz
1.3 % (0.056 dB)
HP 432A power meter & thermistor w/ DC substitution
Absolute RF Power – Measure – 1 µW to 100 mW* (-30 to +20 dBm)
(100 to 300) kHz
>300 kHz to 1 MHz
>1 MHz to 2 GHz
(>2 to 12.4) GHz
(>12.4 to 18) GHz
(>18 to 26.5) GHz
(>26.5 to 40) GHz
(>40 to 50) GHz
1.7 % + 140 nW
1.2 % + 140 nW
1.2 % + 140 nW
1.6 % + 140 nW
2.3 % + 140 nW
2.4 % + 140 nW
2.8 % + 140 nW
4.1 % + 140 nW
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Absolute RF Power – Measure – 100 pW to 1.0 µW (-70 to -30 dBm)
(0.05 to 0.1) GHz
(0.1 to 2) GHz
(>2 to 12.4) GHz
(>12.4 to 18) GHz
(>18 to 34) GHz
(>34 to 40) GHz
(>40 to 50) GHz
2.5 % + 57 pW
1.9 % + 57 pW
1.8 % + 57 pW
2.1 % + 57 pW
2.9 % + 57 pW
3.4 % + 57 pW
5.4 % + 57 pW
8487D power sensor & EPM-series meter
8487D power sensor & EPM-series meter
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
Power meter & thermistor with DC substitution
AC Current – Clamp Meters – (0 to 500) A
(45 to 400) Hz
0.37 % + 0.097 A
Fluke 552XA & 5500A/COIL
AC Current – Clamp Meters – (500 to 1000) A
(45 to 400) Hz
0.38 % + 0.2 A
Fluke 552XA & 5500A/COIL
AC Current – Clamp Meters – Meters – (0 to 500) A
(45 to 440) Hz
0.37 % + 0.1 A
Fluke 5522A w/5500 coil
AC Current – Clamp Meters – Meters – (500 to 1000) A
(45 to 440) Hz
0.39 % + 0.2 A
Fluke 5522A w/5500 coil
AC Current – Clamp on Meter – (10 to 25) A
(45 to 60) Hz
(64 to 440) Hz
0.71 % + 0.03 A
1.3 % + 0.03 A
Fluke 5520A/SC1100, Fluke 5500A/coil
Fluke 5520A/SC1100, Fluke 5500A/coil
AC Current – Clamp on Meter – (150 to 1000) A
(45 to 60) Hz
(64 to 440) Hz
0.74 % + 0.9 A
1.6 % + 0.9 A
Fluke 5520A/SC1100, Fluke 5500A/coil
Fluke 5520A/SC1100, Fluke 5500A/coil
AC Current – Clamp on Meter – (25 to 150) A
(45 to 60) Hz
(64 to 440) Hz
0.71 % + 0.25 A
1.7 % + 0.25 A
Fluke 5520A/SC1100, Fluke 5500A/coil
Fluke 5520A/SC1100, Fluke 5500A/coil
AC Current – Clamp-On Meter – (10 to 25) A
(45 to 65) Hz
(65 to 440) Hz
0.71 % + 0.03 A
1.3 % + 0.03 A
Fluke 5520A/SC600, Fluke 5500A/coil
Fluke 5520A/SC600, Fluke 5500A/coil
AC Current – Clamp-on Meter – (150 to 1000) A
(45 to 65) Hz
(65 to 440) Hz
0.57 % + 1 A
1.0 % + 1 A
Fluke 5520A/SC1100 Fluke 5500A/coil
Fluke 5520A/SC1100 Fluke 5500A/coil
AC Current – Clamp-On Meter – (150 to 1025) A
(45 to 65) Hz
(65 to 440) Hz
0.71 % + 0.9 A
1.6 % + 0.9 A
Fluke 5520A/SC600, Fluke 5500A/coil
Fluke 5520A/SC600, Fluke 5500A/coil
AC Current – Clamp-on Meter – (20 to 150) A
(45 to 65) Hz
(65 to 440) Hz
0.56 % + 290 mA
1.0 % + 290 mA
Fluke 5520A/SC1100 Fluke 5500A/coil
Fluke 5520A/SC1100 Fluke 5500A/coil
AC Current – Clamp-On Meter – (25 to 150) A
(45 to 65) Hz
(65 to 440) Hz
0.73 % + 0.25 A
1.7 % + 0.25 A
Fluke 5520A/SC600, Fluke 5500A/coil
Fluke 5520A/SC600, Fluke 5500A/coil
AC Current – Clamp-on Meter – Non-Toroidal – (150 to 1000) A
(45 to 65) Hz
(65 to 440) Hz
0.67 % + 1 A
1.2 % + 1 A
Fluke 5520A/SC600 Fluke 5500A/coil
Fluke 5520A/SC600 Fluke 5500A/coil
AC Current – Clamp-on Meter – Non-Toroidal – (20 to 150) A
(65 to 440) Hz
1.2 % + 0.29 A
Fluke 5520A/SC600 Fluke 5500A/coil
AC Current – Clamp-on Meter – Toroidal – (150 to 1000) A
(45 to 65) Hz
0.38 % + 0.1 A
Fluke 5520A/SC600 Fluke 5500A/coil
AC Current – Clamp-on Meter – Toroidal – (20 to 150) A
(45 to 65) Hz
(65 to 440) Hz
0.38 % + 0.029 A
0.93 % + 0.032 A
Fluke 5520A/SC600 Fluke 5500A/coil
Fluke 5520A/SC600 Fluke 5500A/coil
AC Current – Generate
(3 to 45) Hz
45 to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
0.020 % + 0.1 mA
0.020 % + 0.05 mA
0.020 % + 0.08 mA
0.19 % + 300 µA
0.39 % + 300 µA
Fluke 5560A
Fluke 5560A
Fluke 5560A
Fluke 5560A
Fluke 5560A
AC Current – Generate – (>0.22 to 2.2) A
(0.02 to 1) kHz
(>1 to 5) kHz
(>5 to 10) kHz
0.027 % + 35 µA
0.046 % + 80 µA
0.7 % + 160 µA
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
AC Current – Generate – (>0.22 to 2.2) mA
(10 to 20) Hz
(>20 to 40) Hz
(>0.04 to 1) kHz
(>1 to 5) kHz
(>5 to 10) kHz
0.025 % + 40 nA
0.017 % + 35 nA
0.013 % + 35 nA
0.020 % + 110 nA
0.11 + 650 nA
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
AC Current – Generate – (>2.2 to 11) A
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.047 % + 0.17 mA
0.095 % + 0.38 mA
0.36 % + 0.75 mA
Fluke 5720A Series II, 5725A
Fluke 5720A Series II, 5725A
Fluke 5720A series II
AC Current – Generate – (>2.2 to 22) mA
(10 to 20) Hz
(>20 to 40) Hz
(>0.04 to 1) kHz
(>1 to 5) kHz
(>5 to 10) kHz
0.027 % + 400 nA
0.018 % + 350 nA
0.014 % + 350 nA
0.021 % + 550 nA
0.11 % + 5.0 µA
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
AC Current – Generate – (>22 to 220) mA
(10 to 20) Hz
(>20 to 40) Hz
(>0.04 to 1) kHz
(>1 to 5) kHz
(>5 to 10) kHz
0.027 % + 4.0 µA
0.018 % + 3.5 µA
0.014 % + 2.5 µA
0.021 % + 3.5 µA
0.11 % + 10 µA
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
AC Current – Generate – (0 to 220) µA
(10 to 20) Hz
(>20 to 40) Hz
(>0.04 to 1) kHz
(>1 to 5) kHz
(>5 to 10) kHz
0.025 % + 16 nA
0.016 % + 10 nA
0.013 % + 8.0 nA
0.028 % + 12 µA
0.11 % + 65 µA
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
Fluke 5720A series II
AC Current – Generate – (0.12 to 1.2) A
(3 to 45) Hz
45 to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.032 % + 0.5 mA
0.025 % + 0.3 mA
0.030 % + 0.3 mA
0.19 % + 0.5 mA
Fluke 5560A
Fluke 5560A
Fluke 5560A
Fluke 5560A
Load More
Results: 1 - 30 of 1862 capabilities